CVE-2017-12724

EUVD-2017-4263
A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The FTP server on the pump contains hardcoded credentials, which are not fully initialized. The FTP server is only accessible if the pump is configured to allow FTP connections.
ProviderTypeBase ScoreAtk. VectorAtk. ComplexityPriv. RequiredVector
NISTPrimary
8.1 HIGH
NETWORK
HIGH
NONE
CVSS:3.0/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H
Base Score
CVSS 3.x
EPSS Score
Percentile: 63%
Affected Products (NVD)
VendorProductVersion
smiths-medicalmedfusion_4000_wireless_syringe_infusion_pump
1.1
smiths-medicalmedfusion_4000_wireless_syringe_infusion_pump
1.5
smiths-medicalmedfusion_4000_wireless_syringe_infusion_pump
1.6
𝑥