CVE-2019-18241

In Philips IntelliBridge EC40 and EC80, IntelliBridge EC40 Hub all versions, and IntelliBridge EC80 Hub all versions, the SSH server running on the affected products is configured to allow weak ciphers. This could enable an unauthorized attacker with access to the network to capture and replay the session and gain unauthorized access to the EC40/80 hub.
ProviderTypeBase ScoreAtk. VectorAtk. ComplexityPriv. RequiredVector
NISTPrimary
6.5 MEDIUM
ADJACENT_NETWORK
LOW
NONE
CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
Base Score
CVSS 3.x
EPSS Score
Percentile: Unknown
Affected Products (NVD)
VendorProductVersion
philipsintellibridge_ec40_firmware
*
philipsintellibridge_ec80_firmware
*
𝑥
= Vulnerable software versions