CVE-2019-5024

A restricted environment escape vulnerability exists in the “kiosk mode” function of Capsule Technologies SmartLinx Neuron 2 medical information collection devices running versions 9.0.3 or lower. A specific series of keyboard inputs can escape the restricted environment, resulting in full administrator access to the underlying operating system. An attacker can connect to the device via USB port with a keyboard or other HID device to trigger this vulnerability.
ProviderTypeBase ScoreAtk. VectorAtk. ComplexityPriv. RequiredVector
NISTPrimary
7.6 HIGH
PHYSICAL
LOW
NONE
CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
Base Score
CVSS 3.x
EPSS Score
Percentile: Unknown
Affected Products (NVD)
VendorProductVersion
capsuletechsmartlinx_neuron_2_firmware
𝑥
≤ 9.0.3
𝑥
= Vulnerable software versions