CVE-2020-12020

EUVD-2020-4336
Baxter ExactaMix EM 2400 Versions 1.10, 1.11, and 1.13 and ExactaMix EM1200 Versions 1.1, 1.2, and 1.4 does not restrict non administrative users from gaining access to the operating system and editing the application startup script. Successful exploitation of this vulnerability may allow an attacker to alter the startup script as the limited-access user.
ProviderTypeBase ScoreAtk. VectorAtk. ComplexityPriv. RequiredVector
NISTPrimary
6.1 MEDIUM
LOCAL
LOW
LOW
CVSS:3.1/AV:L/AC:L/PR:L/UI:N/S:U/C:N/I:L/A:H
Base Score
CVSS 3.x
EPSS Score
Percentile: 17%
Affected Products (NVD)
VendorProductVersion
baxterem2400_firmware
1.10
baxterem2400_firmware
1.11
baxterem2400_firmware
1.13
baxterem1200_firmware
1.1
baxterem1200_firmware
1.2
baxterem1200_firmware
1.4