CVE-2020-12048

EUVD-2020-4364
Phoenix Hemodialysis Delivery System SW 3.36 and 3.40, The Phoenix Hemodialysis device does not support data-in-transit encryption (e.g., TLS/SSL) when transmitting treatment and prescription data on the network between the Phoenix system and the Exalis dialysis data management tool. An attacker with access to the network could observe sensitive treatment and prescription data sent between the Phoenix system and the Exalis tool.
ProviderTypeBase ScoreAtk. VectorAtk. ComplexityPriv. RequiredVector
NISTPrimary
7.5 HIGH
NETWORK
LOW
NONE
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
Base Score
CVSS 3.x
EPSS Score
Percentile: 18%
Affected Products (NVD)
VendorProductVersion
baxterphoenix_x36_firmware
3.36
baxterphoenix_x36_firmware
3.40
𝑥