CVE-2020-25187

EUVD-2020-17878
Medtronic MyCareLink Smart 25000 is 

 vulnerable when an authenticated attacker runs a debug command, which can be sent to the patient reader and cause a heap overflow event within the MCL Smart Patient Reader software stack. The heap overflow could allow an attacker to remotely execute code on the MCL Smart Patient Reader, potentially leading to control of the device
ProviderTypeBase ScoreAtk. VectorAtk. ComplexityPriv. RequiredVector
NISTPrimary
8.8 HIGH
ADJACENT_NETWORK
LOW
NONE
CVSS:3.1/AV:A/AC:L/PR:N/UI:R/S:C/C:H/I:H/A:H
icscertCNA
8.8 HIGH
ADJACENT_NETWORK
LOW
NONE
CVSS:3.1/AV:A/AC:L/PR:N/UI:R/S:C/C:H/I:H/A:H