CVE-2020-27252

Medtronic MyCareLink Smart 25000 is 

vulnerable to a race condition in the MCL Smart Patient Reader software update system, which allows unsigned firmware to be uploaded and executed on the Patient Reader. If exploited, an attacker could remotely execute code on the MCL Smart Patient Reader device, leading to control of the device.
TOCTOU
ProviderTypeBase ScoreAtk. VectorAtk. ComplexityPriv. RequiredVector
NISTNIST
8.8 HIGH
ADJACENT_NETWORK
LOW
NONE
CVSS:3.1/AV:A/AC:L/PR:N/UI:R/S:C/C:H/I:H/A:H
icscertCNA
8.8 HIGH
ADJACENT_NETWORK
LOW
NONE
CVSS:3.1/AV:A/AC:L/PR:N/UI:R/S:C/C:H/I:H/A:H
CVEADP
---
---