CVE-2024-9832

EUVD-2024-50148
There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure.
ProviderTypeBase ScoreAtk. VectorAtk. ComplexityPriv. RequiredVector
NISTPrimary
9.3 CRITICAL
LOCAL
LOW
NONE
CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
Base Score
CVSS 3.x
EPSS Score
Percentile: 25%
Early Detection
Affected products identified ahead of NVD analysis through intelligence sources.
VendorProductVersionSource
baxterlife2000_ventilator_firmware
𝑥
≤ 06.08.00.00
ADP