CVE-2025-24836

With a specially crafted Python script, an attacker could send 
continuous startMeasurement commands over an unencrypted Bluetooth 
connection to the affected device. This would prevent the device from 
connecting to a clinician's app to take patient readings and ostensibly 
flood it with requests, resulting in a denial-of-service condition.
ProviderTypeBase ScoreAtk. VectorAtk. ComplexityPriv. RequiredVector
NISTNIST
7.1 HIGH
ADJACENT_NETWORK
HIGH
NONE
CVSS:3.1/AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:H/A:H
icscertCNA
7.1 HIGH
ADJACENT_NETWORK
HIGH
NONE
CVSS:3.1/AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:H/A:H
CISA-ADPADP
---
---
Awaiting analysis
This vulnerability is currently awaiting analysis.
Base Score
CVSS 3.x
EPSS Score
Percentile: 6%