CVE-2026-105667

EUVD-2026-102895
In the default configuration, the HL7 listener accepts ADT^A40 patient-merge messages without authenticating the sending application. A remote sender can reassign studies and related workflow records from one patient identity to another. The archive's unmerge does not reverse this.
ProviderTypeBase ScoreAtk. VectorAtk. ComplexityPriv. RequiredVector
NISTPrimary
7.5 HIGH
NETWORK
LOW
NONE
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:H/A:N